Description
- 15'' TFT color LCD, multi-language interface(English, Chinese, Russian, Kazakhstan, Czech, Slovak).
- Fanless design,quiet,energy-saving and clean,which reduces the possibility of cross-infection.
- All-round monitor for adult, pediatric and neonate
- With standard interface, big character interface and view bed, convenient to observe.
- Waveform, parameter color and location can be set optionally.
- Display up to 8-channel waveform on one screen.
- Display 7-lead ECG waveform on one screen.
- Adopt digital SpO2 technology,anti-motion and anti-ambient light interference,and measurement can be performed under the circumstance of weak filling.
- NIBP measurement mode: Manual/AUTO/Continuous, storage for 4800-group NIBP data.
- Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
- Drug concentration calculation and titration table functions.
- Connect to Central Monitoring System by 3G, Wi-Fi or wired mode.
- AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
- Anti-high frequency surgical unit, defibrillation-proof(special leads are necessary).
- Touch screen operation, which improves the working efficiency for medical staff.
This device can monitor such parameters as ECG,RESP,SpO2,PR,NIBP and dual-channel TEMP,etc.It integrates parameter measurement module,display and recorder in one device to form a compact and portable equipment. At the same time, its built-in replaceable battery provides convenience for patient moving.
Parameter:
ECG
Lead mode: 3-lead or 5-lead
Lead selection: I, II, III, aVR, aVL, aVF, V
Waveform: 5-lead, 2-channel
3-lead, 1-channel
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s
HR:
Measurement and alarm range: 15~350bpm
Accuracy: ±1% or ±1bpm, whichever is greater
Alarm accuracy: ± 2 bpm
Resolution: 1 bpm
ST-segment monitoring:
Measurement and alarm range: -2.0mV~+2.0mV
Accuracy: -0.8mv~+0.8mv ±0.04mV or ±10%, whichever is greater
Other range: unspecified
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Pacemaker: yes
RESP
Method: R-F(RA-LL) Impedance
Respiration rate:
Measurement and alarm range: 0~150rpm
Resolution: 1 rpm
Measurement accuracy: ±2 rpm
Alarm accuracy: ±3 rpm
Apnea alarm: 10~40s
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
NIBP
Method: Oscillometry
Mode: Manual/AUTO/Continuous
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in Continuous mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean deviation: ±5 mmHg
Maximal standard deviation: 8mmHg
Over-pressure protection:
Adult mode: 297±3 mmHg
Pediatric mode: 240±3 mmHg
Neonatal mode: 147±3 mmHg
SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1%
Measurement accuracy: 70%~100%: ±2%;
0%~69%: unspecified
PR
Measurement and alarm range: 30 ~ 250 bpm
Measurement accuracy: ±2 bpm or ±2%, whichever is greater
TEMP
Channel: dual-channel
Measurement and alarm range: 0 ~ 50℃
Resolution: 0.1 ℃
Accuracy: ±0.1 ℃
EtCO2
Method: Sidestream or Mainstream
Measurement range for CO2: 0~150mmHg
Resolution for CO2:
0.1 mm Hg: 0 ~ 69 mm Hg
0.25 mm Hg: 70 ~ 150 mm Hg
Accuracy for CO2:
0 ~ 40 mm Hg: ±2 mm Hg
41 ~ 70 mm Hg: ±5%
71 ~ 100 mm Hg: ±8%
101 ~ 150 mm Hg: ±10%
AwRR range: 2~150 rpm
AwRR accuracy: ±1 rpm
Apnea alarm: yes
IBP
Channel: dual-channel
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measurement and alarm range: -10~300 mm Hg
Resolution: 1 mmHg
Accuracy: ±2% or 1mm Hg, whichever is greater
Package List:
1×Patient monitor
1×Adult fingertip SpO2 probe (5-pin)
1×Adult NIBP cuff
1×NIBP extension tube
1×ECG lead cable
1×ECG electrode
1×Temperature probe
1×Power cord
1×User Manual
Zahlung & Sicherheit
Ihre Zahlungsinformationen werden sicher verarbeitet. Wir speichern keine Kreditkartendaten und haben keinen Zugriff auf Ihre Kreditkarteninformationen.